6.3.2.
Patient cable
Check the cable regularly for damages and/or bad
electrical contact. We also advise that you keep an
extra patient cable in stock.
6.4. Environment information
Your Endomed 182 contains materials which can be
recycled and/or are noxious for the environment. At
the end of the duration of life, specialised concerns
can take apart the Endomed 182 and sort out the
noxious materials and materials for recycling. By
doing so you contribute to a better environment.
Please ensure that you are well informed of the
local rules and regulations regarding to the re-
moval of equipment and accessories.
22
CHAPTER 7 - FAULT CONDITIONS
7.1. Display's fail to light up
• Check if the mains adapter is connected to the
unit and mains supply.
7.2.
Code ErrX on the display
The unit has discovered a fault using its "self-test".
Remove any applicators or cables from the output
sockets from the equipment, switch the unit back
on. If the code re-appears, contact your supplier.
The unit is probably defective.
* ErrX: number is mentioned on display [6], run-
ning from 1 until and including 9.
7.5. Insufficient or no output current
• Check the patient cable for breaks or poor con-
tact.
• Ensure that the sponges are sufficiently damp.
• If necessary, use a salt solution.
• Clean the electrodes.
• Check for poor contact.