CHAPTER 3 - INDICATIONS AND
CONTRA-INDICATIONS
See also the therapy book that can be ordered sepa-
rately.
3.1. Indications
Diagnostics, electro palpation for:
• pain points;
• trigger points;
• hyperaesthetic regions;
• motor trigger points.
3.2. Therapy
A. Pain control in:
• pain points;
• trigger points;
• hyperaesthetic regions.
B Autonomic disorders such as:
• shoulder/hand syndrome;
• Raynaud's disease;
• Buerger's disease;
• Sudeck's dystrophy;
• neurological diseases;
• myalgia.
C. Muscle stimulation:
• for restoration of kinaesthesia;
• for treatment of atrophy;
• for stimulation of the internal and external sphinc-
ters in the treatment of urinary incontinence;
• revalidation;
• muscle strengthening (sport);
• neurological diseases
• spasm inhibit.
D. Iontophoresis.
3.3. Contra-indications
• pyrexia;
• tumours;
• tuberculosis;
• localised inflammation;
• thrombosis;
• pregnancy;
• pacemakers;
• metal implants.
CHAPTER 4 - INSTALLATION
4.1. Connection
• Mains supply connections must comply with the
national requirements regarding medical rooms.
• Prior to connection of this apparatus to the mains
supply, check that the voltage and frequency
stated on the type plate correspond with the avail-
able mains supply.
• The adapter is a part of the supply circuit on which
the device's safety partly depends. The approv-
als for the Endomed 182 are only valid if used in
combination with this type of adapter ENB-1530.
It is not permitted to connect the Endomed 182
to another type of adapter than type ENB-1530.
4.2. Connection of the mains adapter
• Connect the supplied mains adapter to the con-
nector [14].
• Connect the mains adapter to a wall socket. The
flashing dot in the upper display [3]. indicates
that the unit is connected to the mains supply
and stand-by.
4.3. Switching on and self test
• Switch On the unit, using the On/Off button [1]
• Immediately after switching on, the unit carries
out a self test.
Check that:
• All LED's light up briefly.
• At the end of the test a beep can be heard.
Contact your supplier if this is not the case.
4.4. Disconnecting from the mains
• Switch the Endomed 182 Off, using the button [1].
• Pull out the mains plug of the mains adapter from
the wall socket.
• Disconnect the mains adapter from the Endomed
182.
4.5. Installation
• Do not install the unit in a location near to a heat
source such as a radiator.
• Avoid exposure to direct sunlight, rain, excessive
dust, dampness, mechanical vibrations and
shocks.
• Cordless telephones should not be used in close
proximity of the unit.
• This unit should not be used in so-called "wet-
rooms" (hydrotherapy rooms).
• The unit has to be installed in such way that
liquid cannot enter.
17