Operating Environment - Miniland sweetBeatL Manual

Tabla de contenido
Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 1

7.2. Operating environment

• Temperature: From +5 ºC to +40 ºC
• Humidity: ≤ 80%
• Atmospheric pressure: From 86 kPa to 106 kPa
• Service life: 5 years
7.3. Transportation and storage
• Temperature: From -10 ºC to +55 ºC
• Humidity: <93%
• Atmospheric pressure: From 86 kPa to 106 kPa
7.4. Measurement of the fetal heartbeat
• Sensitivity: ≥ 90 dB
• Measuring range: From 50 to 210 beats per
minute
• Accuracy: ± 2 beats per minute
NORMALIZED SYMBOLS
• The name and the address of the Manufacturer
• Authorized Representative in the European Community
• Batch code
• Attention: Read the operating Instructions for use!
• Applied part of type B
• Disposal in accordance with Directive 2012/19/UE (WEEE)
• Complies with the European Medical Device Directive (93/42/EEC)
and amended by directive 2007/47/EC requirements.
Notified body: MEDCERT GmbH (CE0482)
• Serial number
8. IDENTIFYING AND SOLVING PROBLEMS
Problem
Nothing happens when
you try to turn the
apparatus on
You mobile device is
unable to detect the
heartbeat monitor in
its range of coverage
when conducting the
Bluetooth device search
Possible cause
Flat battery
The heartbeat monitor is not
turned on
Your mobile Bluetooth system
is not turned on
The heartbeat monitor battery
has gone flat
You mobile device is not
compatible with Bluetooth
BLE 4.0
sweetBeat is out of the range
of your mobile device´s
Bluetooth
7.5. Electrical specifications
• Battery: Rechargeable 3.7 V Ion-Lithium
7.6. Specifications of the probe
• Product name: Fetal Doppler
• Model: iFM-10B
• Type: Continuous Doppler
• Water resistance: IPX1
• Acoustic operating frequency: 2.5 MHz ± 15%
• Ultrasonic output power: ≤ 20 mW
• Effective emission area: 245 mm² ± 15%
• p_ < 1MPa
• I
< 20 mW/cm²
ob
• I
< 100 mW/cm²
spta
Solution
Charge the heartbeat monitor as
illustrated in item 5.1.
Turn it on as illustrated in item 5.3.
Turn on your mobile Bluetooth system
as illustrated in item 5.3
Charge the heartbeat monitor as
illustrated in item 5.1.
Check sweetBeat compatibility in
section 7.
TECHNICAL SPECIFICATIONS
Move your mobile device closer to the
heartbeat monitor
15
ENGLISH
Tabla de contenido
loading

Este manual también es adecuado para:

89304

Tabla de contenido