Lionelo care Nebi Air Mask Manual De Usuario página 6

Tabla de contenido
Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 36
and other accessories and then wipe the residues away with a sterile gauze. Do not touch the mesh disc with your hand or sharp objects due to the risk of damage.
Cleaning
Before cleaning, switch the device off and disconnect all the accessories. The device must not be connected to the mains.
Immerse the accessories in warm water (no more than 40°C) for ca. 5 minutes.
After cleaning, dry the accessories with a sterile gauze.
Store the accessories in a dry and clean place.
The main unit should be cleaned with a dry cloth. If there are any medication residues on the device, they should be cleaned with a damp, sterile gauze and left
to dry completely.
Please note! Do not immerse the mask in hot water.
Disinfection
After every use, disinfect the medicine cup, spray nozzle, mask, cup cover and mouthpiece.
1.
Disinfecting with hydrogen peroxide
Immerse the accessories in 2% hydrogen peroxide for ca. 10 minutes. After you have disinfected them, rinse the accessories with water and dry with a sterile gauze
or leave to dry. Do not use any strong oxidizing agents.
2.
Disinfecting with ethyl alcohol
Immerse the accessories in ethyl alcohol for ca. 10 minutes. After you have disinfected them, rinse the accessories with water and dry with a sterile gauze or leave to dry.
Drying
Shake the medicine cup to remove any remaining water from the cup and the mesh disc.
Leave the accessories to dry for at least 4 hours.
Ensure that all the accessories are perfectly dry, including but not limited to the mesh disc and medicine cup.
Storage and maintenance
1.
Storage conditions
Ambient temperature: -10°C~+50°C
Other: non-corrosive gases, good ventilation, avoiding high temperature, humidity and direct sun rays.
2.
Storage instructions
The device has a shelf life of 5 years, provided the above storage conditions are ensured. The medicine cup should be replaced after no more than 18 months.
After use, clean and disinfect the device and the accessories. Store the device in the original packaging. Avoid any shock.
3.
Operating conditions
Temperature: 5°C~40°C
Relative humidity, non-condensing: 30-80% RH
Power supply: lithium battery 3.7 V DC 230 mAh
Troubleshooting
Problem
The device does not switch on.
Poor nebulization.
The device switches off or makes a strange noise.
Specific indications for children
The mask should cover the mouth and the nose.
Do not nebulize a sleeping child.
Please note! Children require adult help and supervision when using the device.
Electromagnetic compatibility (EMC) symbols
This device fulfills the requirements relating to the electromagnetic compatibility (IEC60601-1-2).
Use the device according to the provided EMC information.
The device must not be subject to any interference from radio communication devices. Such devices should be used at least 30 m away from it.
Do not use the device near any devices generating strong magnetic field and, whenever this is required, the device should be watched to check its correct operation
in the relevant configuration.
Using any accessories and cables other than provided may result in increased emissions or reduced product resistance.
Relative humidity, non-condensing: 30–85% RH
Power supply parameters (not included): 5.0 V DC 1.0 A
Ambient pressure: 86–106 kPa
Ambient pressure: 86–106 kPa
Cause/solution
Check the battery level.
Check if the medicine cup is filled.
Keep the device upright.
Check if the cup is filled with an appropriate, water-soluble medicine.
Check if there is sufficient amount of the medicine.
Check if the mesh disc is not clogged.
Check if there is sufficient amount of the medicine.
Tabla de contenido
loading

Tabla de contenido