Nexo Rose 202 2SH Manual página 93

Tabla de contenido
Idiomas disponibles
  • ES

Idiomas disponibles

2) Contra angle head part
Screw the clasp on the straight handpiece until it cannot move, and take out the head of the contra angle handpiece.
Insert the turbine lubricant in the hole as Fig 15 shown. Lubricant one time each day, using the turbine lubricant 1-2
drop.
CAUTION
Improper maintenance and care can lead to premature wear and malfunction. We recommend using original NEXO
care products
9 > STERILIZATION
1.
Suitable for both steam. Autoclaving procedure:
2.
3.
Wipe off dirt with alcohol-soaked cloth according to the manual item 7 and item 8.
4.
Lubricate with SPRAY. Operate the handpiece for a few seconds after lubrication with Air Turbine Oil (Ref. 6.
LUBRICATION).
5.
Place into autoclaving pouch and seal it in accordance with instructions on the pouch,
6.
Autoclave the handpiece for 20 min. at 120ºC, or 15 min, at 135ºC.
7.
Please dry the handpiece by dryer for 134ºC (15minutes) after high temperature sterilization and disinfection.
93