Compatibilidad Electromagné t ica (CEM)
Datos generales CEM
Emisión
RF-Emission according EN 55011
RF-Emission according EN 55011
Generation of mains harmonics
following IEC 6100-3-2
Generation of mains harmonics
following IEC 6100-3-3
Tabla CEM 202
Susceptibility
ESD
IEC 61000-4-2
Bursts
IEC 61000-4-4
Surges
IEC 61000-4-5
Voltage drops etc.
IEC 61000-4-11
H-field at 50/60Hz
IEC 61000-4-8
Tabla CEM 204
Susceptibility
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
Tabla CEM 206
Output power of transmitter W
0,01
0,1
1
10
100
Sujeto a modifcaciones técnicas
IEC 60601-test level
±6kV cd
±8kV ad
±2kV mains
±1kV I/O
±1kV dm
±2kV cm
Reduction to
5% for 10ms/ positive amplitude
5% for 10ms/ negative amplitude
40% for 100ms
30% for 500ms
0% for 5000ms
3A/m
IEC 60601-test level
3Veff
150kHz to 80 MHz
3V/mf
80 MHz to 2,5 GHz
Safety distance depending on frequency/m
150kHz to 80MHz 80MHz to 800MHz 800MHz to 2,5 GHz
0,12m
0,12m
0,37m
0,37m
1,17m
1,17m
3,69m
3,69m
11,67m
11,67m
Conformidad
grupo 1
clase B
Battery operation, not applicable
Battery operation, not applicable
actual level
±2kV,±4kV ±6kV cd
±8kV ad
Battery operation, not applicable
Battery operation, not applicable
Reduction to
Battery operation, not applicable
Battery operation, not applicable
Battery operation, not applicable
Battery operation, not applicable
Battery operation, not applicable
No magnetically sensitive sensors or
parts, not applicable
actual level
3Veff
3V/m
0,24m
0,74m
2,34m
7,38m
23,34m
Liftkar PT I 20