Maintenance - Blatchford BMK2 Knee Instrucciones De Uso

Ocultar thumbs Ver también para BMK2 Knee:
Tabla de contenido
Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 74

5 Maintenance

Maintenance must be carried out by competent personnel.
It is recommended that the following maintenance is carried out annually:
• Check all non-adjuster screws for tightness (see Construction section) (DO NOT tighten the
covered nut at the distal end).
• Clean the brake pad and replace the extension stops (see Section 10 Maintenance of
Stabilizing Mechanism for procedure).
• In the event of unusual noises suchs as squeaks and creaks and/or reduced stability, carry
out maintenance of the stabilizing mechanism, see Section 10 Maintenance of Stabilizing
Mechanism. If noises persist contact your Blatchford representative.
• Wear and tear on the extension buffer is acceptable. If excessive noise is present at full
extension replace the extension buffer. See Section11 Maintenance of Extension Buffer.
• Check for excessive Medio-Lateral play of the chassis.
• If the trailing edge of the brake band pad is worn down to the plastic of the brake band,
replace the complete brake band. (Note: Individual pads cannot be replaced.)
• Ensure orientation of brakeband is correct on reassembly. Always check stability before
using.
The user must be handed the User Guide supplied, and be advised to report the following to the
practitioner:
• Any changes in body weight or activity level
• Any changes in performance of this device
Changes in performance may include:
• Restricted swing
• Changes in smoothness of swing
• Reduced stability during stance
• Any unusual noise
Cleaning
Use a damp cloth and mild soap to clean outside surfaces, DO NOT use aggressive cleansers.
Thoroughly dry before use.
6
938268PK1/1-0422
Tabla de contenido

Solución de problemas

loading

Este manual también es adecuado para:

Bmk2

Tabla de contenido