ROUTINE MAINTENANCE AND
REPLACEMENT PARTS
IMPORTANT:
A listing of the Safety Precautions to be observed
can be found in S
this information.
5.1 Routine
Maintenance
WARNING – PERSONAL INJURY
AND/OR EQUIPMENT DAMAGE
HAZARD:
• Regularly scheduled preventive
maintenance is required for
safe and reliable operation of
this
equipment.
STERIS to schedule preventive
maintenance or obtain the
necessary washer/disinfector
Maintenance
preventive
maintenance
done by the Customer.
• Repairs and adjustments to
this equipment must be made
only by STERIS or STERIS-
trained
service
Repairs
and
performed
by
personnel or installation of
unauthorized
cause personal injury, result
in
improper
performance,
warranty or result in costly
damage.
Contact
regarding service options.
CAUTION – POSSIBLE EQUIPMENT
DAMAGE:
Use
cleaners when cleaning Long
Three-Level MIS Rack. Follow
directions on containers and rub in
a back-and forth motion (in same
direction
as
surface
Abrasive cleaners will damage
stainless steel.

Routine Maintenance and Replacement Parts

1 of this manual. Do not use the equipment until you have become familiar with
ECTION
NOTE: The Three-level Ophthalmic Rack with Removable Upper
Insert – Long must be inspected regularly to ensure no sediment has
accumulated in rack components. As the usage of this rack can be
variable, it is recommended that the washer spray arm be maintained
every week. This frequency represents approximately every
40 cycles.
1. Inspect all hollow jets, connections, caps, adapters, spray jets,
Contact
2. Inspect all rubber caps and silicone tubings. Replace if torn,
Manual
if
is
Inspect the lower large mesh basket as follows:
1. Remove socket screw at basket bottom.
2. Use compressed air or a small brush to clean.
personnel.
3. Replace socket screw.
adjustments
unqualified
parts
could
1. Clean first rotary spray arm assembly as follows:
equipment
void
the
STERIS
2. Remove and clean second rotary spray arm assembly located in
nonabrasive
1. Rotate counter clockwise and pull to remove each end spray
grain).
2. Use a fine wire to gently remove sediment from spray nozzles.
Operator Manual
Hollow Jets, Luer locks and Adapters:
spindles and nozzles. Unclog, if necessary, by using a fine wire
(approximately wire gauge of a paper clip).
discolored or damaged.
Lower Mesh Basket:
Rotary Spray Arm Removal
a. Remove rotary spray arm assembly.
b. Use a fine wire (approximately wire gauge of a paper clip) to
clean out sediment from spray nozzles.
c. Rinse under running water.
d. Reinstall rotary spray arm assembly.
lower section of rack as explained in Step 1.
Rack Corner Spray Nozzles Removal and Cleaning
nozzle.
Rinse under running water.
when using and servicing this equipment
5
5-1
10095339