preparation and even damage the tooth. Make sure that
the spray water is at least 50ml/min at a pressure of 2 bar
and that the spray air is at least 1.5Nl/min at a pressure
of 2 bar.
WARNING!
Do not use the handpiece to move the patient's cheek
while the device is functioning!
WARNING!
Do not look at the LED light for a long time when it is
illuminated.
WARNING!
Always check the handpiece for damage before using it.
WARNING!
Make sure that the micromotor in which the handpiece
will be inserted has a maximum rotation of 40,000 rpm.
30
Eng
Before use, make sure that all the o-rings are present on
the micromotor coupling.
WARNING!
It can be dangerous to use the handpiece without a
speed and direction controller.
- The dental unit with which this handpiece is used must
be equipped with a speed and direction controller.
- For reasons of legal responsibility for safety, reliability and
performance, a statement to this effect must also appear in
the documentation provided with the dental unit.
DANGER!
Always wear gloves to avoid the risk of lesions and
infections
when
checking,
extracting teeth.
CAUTION!
Use of unfiltered, hard water can lead to blockages and
premature wear in hoses, fittings and spray heads.
cleaning,
repairing
or