Holster Specifications
Electrical Conformity
Fluid Ingress
Flammability
Dimensions
AC Power Cord and Adapters

System Specifications

Classification
Environmental Conditions
Mode of Operation
Electromagnetic Compatibility (EMC), Guidance, and Manufacturer Declarations
The Mammotome® elite Biopsy System requires special precautions regarding electromagnetic compatibility (EMC)
and must be installed and used in accordance with the EMC information provided in this installation guide. The
Mammotome® elite Biopsy System is intended for use in the electromagnetic environment specified below. The
customer or user of the Mammotome® elite Biopsy System should assure that is used in such an environment.
Guidance and manufacturer's declaration – electromagnetic emissions
The Mammotome® elite Biopsy System is intended for use in electromagnetic environments specified below. The customer or user of the
Mammotome® elite Biopsy System should assure that is used in such an environment.
Emission Test
RF Emissions CISPR 11
RF Emissions CISPR 11
Harmonic Emissions
IEC 61000-3-2
Voltage fluctuations/flicker
emissions
IEC 61000-3-3
Guidance and manufacturer's declaration – electromagnetic immunity
This medical equipment has passed all required testing for electric shock, fire and mechanical
hazards in accordance with UL 60601-1, IEC/EN 60601-1, CAN/CSA C22.2 No 601-1
Not protected against the ingress of fluids
Equipment not suitable for use in the presence of flammable anesthetic
Length: 18.9 cm
Width: 5.0 cm
Height: 3.8 cm
Use only with supplied AC power accessories:
• Model : FRM015-S09-1 p/n TRM15-S10-I-351359
• Input Voltage: 100-240VAC
• Input Current: 0.7A-0.4A
• Input Frequency: 50-60Hz
• Output Voltage:10 Volts DC
• Output Current: 1.5 A
• Max Watts: 15W
Class II, internally powered Type BF equipment
Transportation and Storage:
• -35°C to +54°C (-31°F to 129°F)
• 10% to 95% Relative Humidity, Non-Condensing
• 500 to 1060 hPa Atmosphere Pressure
Operation:
• +5°C to +40°C (41°F to 104°F)
Continuous
Compliance
Guidance
Group 1
The Mammotome® elite Biopsy System uses RF energy only for its
internal function. Therefore, its RF emissions are very low and are not
likely to cause any interference in nearby electronic equipment.
Class A
Not Applicable
The Mammotome® elite Biopsy System is suitable for use in all
establishments other than domestic and those directly connected to
the public low-voltage power supply network that supplies buildings
Not Applicable
used for domestic purposes.