Starting A Verification - Mettler Toledo SevenDirect SD30 Manual De Usuario

Tabla de contenido
Idiomas disponibles

Idiomas disponibles

To ensure the most accurate readings, you should perform calibrations regularly.
Running Direct Calibration
Ensure that the appropriate buffers/standards have been selected.
1 Select an appropriate method according to [Selecting a method   Page 14].
2 Select a sensor if necessary according to [Selecting a sensor   Page 14].
3 Tap Calibration.
4 Tap Start calibration.
5 Enter the lot number if Methods > Calibration > Check standard lot number is active and confirm with
OK.
The sensor name, the endpoint type and buffers/standards points are displayed.
6 Place the sensor in a calibration buffer/standard.
7 Tap Read.
The font of the measurement values turns to light blue.
8 The measurement stops when the endpoint criterion is fulfilled. See [Endpoint criterion   Page 16].
appears in front of the passed calibration point.
9 Rinse the sensor with deionized water and place the sensor in the next calibration buffer/standard.
10 Tap Read.
The font of the measurement values turns to light blue.
11 The measurement stops when the endpoint criterion is fulfilled. See [Endpoint criterion   Page 16].
appears in front of the passed calibration point.
12 Rinse the sensor with deionized water and repeat the steps with all buffers/standards.
A window pops up with calibration data.
13 Tap OK (Adjust) to save the result.
- or -
Tap Reject to reject the calibration and return to Home screen.
Finalize a Calibration in Advance
If Home screen > Menu > Settings > Analysis settings > Finalize pH calibration at any point is active, a
button Calculate will display for the multi-point pH calibration. You can tap it to finalize the calibration in
advance.
To finalize the calibration in advance, proceed as follows:
Analysis settings is active.
The calibration is a multi-point pH calibration.
At least one calibration point of a multi-point calibration is passed.
1 Tap Calculate.
2 Tap OK (Adjust) to save the result.
- or -
Tap Reject to reject the calibration and return to Home screen.
The meter will do calibrations in the user-defined order in the active method. If one calibration point fails, you
should redo this calibration point and cannot skip it.

5.4 Starting a verification

The latest verification data will be stored into the data of the active sensor.
Ensure that the appropriate buffers/standards have been selected.
1 Select an appropriate method according to [Selecting a method   Page 14].
2 Select a sensor if necessary according to [Selecting a sensor   Page 14].
3 Tap Verification.
4 Tap Start verification.
SevenDirect™ SD30
15
Tabla de contenido
loading

Tabla de contenido