Press the Middle Control button below the Start
Therapy icon to start the therapy. The LEDs on the
front and underside of the pump change to green.
NOTE
If the pump is switched on and one or more garments
are connected to the pump, but therapy is not started
within 15 minutes, an alarm will occur. If no action is
taken to start therapy or silence the alarm, then 15
minutes after the alarm starts, the pump will go into an
"Off" state (refer to "Troubleshooting" on page 25).
It is recommended that the following checks are carried
out at the start of, and throughout, the therapy:
• Check the LCD screen icons to confirm that the
correct type of garment(s) have been connected.
• During garment inflation, check the LCD screen to
confirm that there are no fault indicators displayed
and that the correct pressure is being supplied. The
default target inflation pressures for the different
garments are:
• Foot garment: 130 mmHg.
• Uniform calf and calf & thigh garments:
40 mmHg.
• Sequential calf and calf & thigh garments:
45 mmHg.
• Check that there are no kinks in the tubeset.
• Check that the tubeset and connectors do not cause
the patient any discomfort.
• Regularly check that the garments remain correctly
fitted to the patient.
The pump will inflate and deflate each garment in turn,
as follows, starting with the garment connected to
tubeset connector 1. The garment pressure is
displayed in the top left side of the screen. The Therapy
Time is displayed in the bottom right of the screen.
NOTE
The following example shows two foot garments fitted
to the pump.
1. Both garments are initially deflated, and the
garment icons are shown as outlines.