36.8
High temperature point calibration:
For example:
Before calibration, the screen displays 60°C but the actual liquid temperature is
65.0°C;
After calibration, the screen displays 64.7°C and the actual liquid temperature is
65.0°C;
Operation
Display
°C
HCP
60.0
65.0
SET
64.7
The setting will exit automatically after 5S without
operation
Leave the bath to stabilize and check the calibration
Select the menu option
Select the high temperature point calibration
setting menu option
Press set to enter the calibration value
Enter the reference thermometer value
Press set to store the calibration value
The setting will exit automatically after 5S without
operation
Leave the bath to stabilize and check the calibration
12