battery (when the scale is connected to AC power, the system will charge the
storage battery automatically whether the scale is turned on or off. It is advised
to charge the electronic scale after it is turned off).
9. PRECAUTION
Never put the electronic scale in the rain or flush it with water. Particularly the
inner parts of the scale should avoid water.
It is forbidden to put the scale in the place with high temperature or humidity.
Avoid any strong shock or heavy weight. Don't put any article or give pressure
over maximum capacity on the scale pan even the scale is not in use.
Wrong wires connection of storage battery is forbidden.
Right connection: Red wire should be connected with red end; Black wire should
be connected with black end.

10. TROUBLESHOOTING

Failure
Description
AC indicator lamp
Failure to connect
doesn't light up
with the AC power
EN
While using stor-
No DC power in
age battery, no
the scale
display and sound
when turning on
Display "—Lb—"
Insufficient of
with alarm sound
voltage of battery
and turn itself off
automatically
Unstable weighing
Insufficient of
display
voltage or get
damped
Possible reason
Solution
1. Poor contact
1. Insert the plug
with the power
well or change
plug
socket
2. Broken power
2. Change power
cord
cord
3. Blown fuse
3. Change fuse of
the same type
1. Damaged stor-
1. Change the
age battery, no
storage battery of
voltage
the same type
2. Wire's poor
2. Contact wire
contact with stor-
with battery well
age battery
(note: color cor-
responding)
Too long use of
Charge the scale
battery
by connecting
with the AC power
supply
Continue to use
Charge the scale
the battery when
for 10 hours by
the scale gives
connecting with
alarm sound of in-
the AC power
sufficient voltage
supply or clean
or the environ-
the main board
mental humidity
and dry it
is too high
16
16
Rev. 06.04.2022
11. CALIBRATION METHOD
1.
Turn on the scale.
2.
While self-checking is over, the indicator will display "0.000"
3.
Input code "52411"or "2013", and press key [TARE] to enter into calibration
mode.
4.
Press key [1] to set capacity. For example, if the capacity is 30kg,input
"30000". Press key [ADD] to confirm.
5.
Press key [2] to set division, press any number key, it will display the division
of "1" "2" "5" "10", choose the division you want. Press key [ADD] to confirm.
6.
Press key [4] to set the position of point. Press any number key to choose the
position, then press key [ADD] to confirm.
7.
Press key [5] to set the function of round, press any number key to choose
the function of round is ON or OFF. Then press key [ADD] to confirm.
8.
Press key [6] to set the keyboard is 14 keys or 24 keys. Press any number key
it will display "ON" or "OFF"(ON means 24 keys, OFF means 14keys). Then
press key [ADD] to confirm.
9.
Press key [7] to choose the point's number, press any number key it will
display: "0" "1" "2"("0" means unit and total window will show "0", "1" means
unit and total window will show "0.0", "2" means unit and total window will
show "0.00"). Then press key [ADD] to confirm.
10. Input the value of poise, press key [ADD], then put on the poise, press[ZERO]
to end calibration.
SAFE REMOVAL OF BATTERIES AND RECHARGEABLE BATTERIES
Worn batteries must be removed from the device proceeding analogously to their
installation. Recycle batteries.
DISPOSING OF USED DEVICES Do not dispose of this device to municipal waste
systems. Hand it over to electric and electrical device recycling and collection
point. Check the symbol on the product, user manual and packaging. The plastics
used to construct the device can be recycled in accordance with their markings.
By recycling you are making a significant contribution to the protection of our
environment. Contact local authorities for information on your local recycling
facility.