ELECTROMAGNETIC COMPATIBILITY INFORMATION

Table 1
Guidance and manufacturer's declaration – electromagnetic emissions
The GCE834 is intended for use in the electromagnetic environment specified below. The customer or the user of the GCE834 should assure
that it is used in such an environment
Emission test
Emissions RF
CISPR 11
RF Emissions
CISPR 11
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations/flicker emissions
IEC 61000-3-3
Table 2
Guidance and manufacturer's declaration – electromagnetic immunity
The GCE834 is intended for use in the electromagnetic environment specified below. The customer or the user of the GCE834 should assure
that it is used in such an environment.
Immunity test
Electrostatic discharges
IEC 61000-4-2
Electrical fast transient/burst
IEC 61000-4-4
Surge
IEC 61000-4-5
Voltage dips, short interrup-
tions and voltage variations
on power supply input lines
IEC 61000-4-11
Power frequency (50/60Hz)
magnetic field
IEC 61000-4-8
NOTE: U
is the a.c. mains voltage prior to application of the test level.
T
Compliance
Electromagnetic environment – guidance
Group 1
The GCE834 uses RF energy only for its internal function. Therefore, its RF emis-
sions are very low and are not likely to cause any interference in nearby electronic
equipment.
Class B
The GCE834 is suitable for use in all establishments including domestic and those
directly connected to the public low-voltage power supply network that supplies
buildings used for domestic purposes.
Class A
Complies
IEC 60601 Test level
±8 kV contact
±15 kV air
Power supply
lines :±2 kV
input/output lines:±1 kV
line(s) to line(s): ±1 kV
line(s) to earth: ±2 kV
100 kHz repetition frequency
0% 0.5 cycle
At 0º, 45 º, 90 º, 135 º, 180 º,
225 º, 270 º and 315 º
0% 1 cycle
And
70% 25/30 cycles
Single phase: at 0
0% 300 cycle
30 A/m
50Hz/60Hz
Compliance level
±8 kV contact
±15 kV air
Power supply
lines :±2 kV
line(s) to line(s): ±1 kV
0% 0.5 cycle
At 0º, 45 º, 90 º, 135 º, 180 º,
225 º, 270 º and 315 º
0% 1 cycle
And
70% 25/30 cycles
Single phase: at 0
0% 300 cycle
30 A/m
50Hz/60Hz
49
Electromagnetic environment – guidance
Floors should be wood, concrete or ceramic tile.
If floors are covered with synthetic material, the
relative humidity should be at least 30 %.
Mains power quality should be that of a typical
commercial or hospital environment.
Mains power quality should be that of a typical
commercial or hospital environment.
Mains power quality should be that of a typical
commercial or hospital environment.
Power frequency magnetic fields should be at
levels characteristic of a typical location in a typi-
cal commercial or hospital environment.