• Connect the compressed air hose [8] to the bottom of
the medication container [10].
• Connect the other end of the compressed air hose [8] to
the nebuliser hose connection [6] by turning it slightly.
Charging the nebuliser battery
To charge the nebuliser battery, proceed as follows:
• Plug the micro USB cable included in delivery into the
micro USB port on the nebuliser.
• Plug the other end of the micro USB cable into the USB
port on the mains adapter and insert the mains adapter
into the socket.
Note
During the charging procedure, between one (battery al-
most empty) and four (battery fully charged) LEDs on the
battery status indicator [2] flash according to the battery
charge status. As soon as the battery is fully charged, all
four LEDs are permanently illuminated.
Switching on the nebuliser
To turn on the nebuliser, proceed as follows:
• Press the ON/OFF button on the nebuliser. The ON/
OFF button illuminates blue. The nebuliser is now op-
erational. The battery must be sufficiently charged for
the nebuliser to function correctly.
Note
The blue LEDs on the battery status indicator show the
current battery state.
1 blue LED flashes
1 blue LED lights up = Battery capacity 25%
2 blue LEDs light up = Battery capacity 50%
3 blue LEDs light up = Battery capacity 75%
4 blue LEDs light up = Battery capacity 100%

5. Operation

Important
• For hygiene reasons, it is essential to clean the atomiser
[9] and the accessories after each treatment and to
disinfect them after the last treatment of the day.
• The accessories may only be used by one person; use
by several people is not recommended.
• If the treatment involves inhaling several different med-
ications one after the other, please be aware that the
atomiser [9] must be rinsed under warm tap water fol-
lowing every usage. See "7. Cleaning and disinfection".
• Please observe the notes on changing the filter in these
instructions for use.
• Check that hose connectors are firmly attached to the
nebuliser [6] and the atomiser [9] before each use of the
device.
• Before use, check the device is working correctly by
briefly switching on the nebuliser (together with the con-
nected atomiser, but without medication). If air comes
out of the atomiser [9], the device is working.
5.1 Inserting the atomiser insert
• Open the atomiser [9] by twisting the top anticlockwise
against the medication container [10]. Place the atom-
iser insert [11] into the medication container [10].
18
= Battery is almost flat, nebu-
liser will soon switch off.