Program: TOL
Afterwards you will return to the SETUP ‐ Menu.
Quitting the SETUP – Menu
2x short
For confirming and saving.
The instrument returns to the last program used
1x long
„ERR 06" in the display could be caused by the following:
• Tolerance limit is out of the acceptable range
• The lower tolerance limit is above the upper tolerance limit.
• The upper tolerance limit is below the lower tolerance limit.
Activating / Deactivating the tolerance marks
Activating TOL – Program
(Example: MIN+TOL – Program)
1x long
Deactivating TOL – Program
(Example: MIN – Program)
1x long
Example: 10,00 ± 0,1mm
Measuring result out of tolerance:
pointer
The LED goes of within 5s after finding the measuring value and lights up again after
finding a new measuring value.
12
Tolerance marks (‐)
LED red
Tolerance
marks
Measuring result within tolerance:
LED green
Toleranc marks
(‐)
Pointer
Tolerance
marks (+)