Trouble-Shooting - diabetikerbedarf adia Instrucciones De Uso

Tabla de contenido
Idiomas disponibles

Idiomas disponibles

8. Trouble-shooting

The following chart may help you identify certain problems, but may not
solve all the problems that could occur. Contact your authorized representative
or the manufacturer if the problem continues.
Message
What it means
Problem with the
blood glucose test
meter
Caused by either used
or wet blood glucose
test strip
Caused by less blood
or irregular blood fill
Caused by reagent
damage in the test
strip
User applied the
blood sample before
the LCD shows the
blinking
.
Action required
Place the battery again,
and set the blood glucose
test meter. If the problem
persists, please contact
your authorized represen-
tative or the manufacturer.
Please insert a new glucose
test strip and perform your
test again.
Please insert a new glucose
test strip and apply the
blood sample.
Please insert a new glucose
test strip and perform your
test again.
Wait for the meter to display
the blinking icon before
applying your blood sample.

Capítulos

Tabla de contenido
loading

Tabla de contenido