The Glucose control solution should be used
Before using a new box of test strips.
Whenever you suspect the blood glucose test meter or blood glucose test
strip is not functioning properly.
If your blood glucose test results are not consistent with your symptoms,
or if you think results are not accurate.
If you have dropped the blood glucose test meter.
For quality control in the point of care usage.
For teaching or learning the system.
When using the meter for blood glucose tests after any disinfection
procedure.
To check your blood glucose meter please use the control solution in the
concentration (low, medium, high) that is next to your measured values.
For the range of each control solution please refer to the label on the strip vial.
Firmly insert the test strip into the meter test port. Insert
deeply in direction of arrow on test strip (arrow up). Please
do not insert the glucose test strip upside down.
When you insert the glucose test strip into the
glucose test meter, the power automatically adjusts
the meter accordingly.
Press and release either the ▲ or ▼ button, then,
Control Solution bottle icon will appear.
Then, press
.

Capítulos