Temperature Settings
1. Press
to enter setup window. Scroll down by pressing
set. The display shows Set °C/°F, Temp Cal and Temp Coeff.
2. Press
to select Set °C/°F. Scroll down by pressing
3. Press
to select temperature unit. The display shows the selected temperature setting with a
Key Info
Scroll
Accept
EXIT
Temperature Calibration
1. Press
to enter setup window. Scroll down by pressing
2. Press
to select Temp Set. The display shows Set °C/°F, Temp Cal and Temp Coeff.
3. Scroll down by pressing
Temp Cal.
4. The lower display shows the current measured temperature reading based on the last set offset and the upper
display shows the current measured temperature reading based on factory default calibration.
5. Dip the tester into a solution of known temperature and allow time for the in built temperature sensor to stabilize.
6. Press
to adjust the temperature value or press the
value of the solution.
Note: To exit this program without confirming the calibration, press
Key Info
Scroll
Accept
EXIT
Temperature Coefficient
1. Press
to enter setup window. Scroll down by pressing
2. Press
to select Temp Set. The display shows Set °C/°F, Temp Cal and Temp Coeff.
3. Scroll down by pressing
4. Press
to select Temp Coeff or
5. Press
to confirm the Temp Coeff value. The new value is automatically confirmed with a
Key Info
Scroll
Accept
EXIT
4
All manuals and user guides at all-guides.com
Measure
Set °C/°F
Settings
Temp Cal
Temp Set
Temp Coeff
Reset
to toggle between Set °C/°F, Temp Cal and Temp Coeff. Press
Measure
Set °C/°F
Settings
Temp Cal
Temp Set
Temp Coeff
Reset
to toggle between Set °C/°F, Temp Cal and Temp Coeff.
to adjust the Temp Coeff.
Measure
Set °C/°F
Settings
Temp Cal
Temp Set
Temp Coeff
Reset
to select Temp Set. Press
to toggle between °C and °F.
°C
°C
°F
to select Temp Set.
to confirm the calibrated value as the new temperature
.
Temp Cal
23.4°C
23.4 °C
23.4 °C
to select Temp Set.
Adjust
2.00
Temp Coeff
2.00
Elite PCTS Tester Operating Instructions
enter to select Temp
.
to select
.