AC Measurement
AC measurements are usually displayed as RMS (root mean squared) values. Two methods of
AC measurement are average-responding RMS calibrated and true RMS-reading.
The average-responding RMS calibrated method takes the average value of the input signal,
multiplies it by 1.11, and displays the result. This method is accurate if the input signal is a
pure sine wave.
The true RMS-reading method uses internal circuitry to read the true RMS value. This method
is accurate, within the specified crest factor limitations, whether the input signal is a pure sine
wave, a square wave, sawtooth, half wave or signal with harmonics. The ability to read true
RMS provides much more measurement versatility. The Greenlee CMP-200 is a true RMS
meter.
The Waveforms and Crest Factors table shows some typical AC signals and their RMS values.