WARNING! Risk of injury! Do not activate the electrosurgical unit
while you are connecting the cable.
Connect the instrument to the cable. Then plug the connector on
the cable into the MF socket on the VIO electrosurgical unit.
20196-115, 20196-118, 20196-119: These products feature
electronic instrument detection. That means that as soon as the
product is plugged into the MF socket, the VIO electrosurgical
unit recognizes it and sets the standard values for mode and ef-
fect automatically. The set modes cannot be changed. However,
you can change the standard values for effects within a certain
range.
6
Cleaning, disinfection, sterilization
6.1
Safety instructions for processing
May be cleaned in an ultrasonic bath.
On no account use sharp or abrasive objects for cleaning purpos-
es.
Disinfectants must be rinsed off thoroughly after use.
Maximum air pressure during drying: 29 psi (2 bar).
Maximum water pressure during rinsing: 29 psi (2 bar).
During machine cleaning and disinfection, do not exceed 203 °F
(95 °C).
Observe the specifications from the manufacturer of the deter-
gent and disinfectant and in the chapter "Overview of Validated
Procedures".
During sterilization, do not exceed 280 °F (138 °C).
Do not sterilize in hot air oven.