Test With Dummy Load; Pump Over-Temperature Test; Functional Test - Varian Turbo-V70 Manual De Instrucciones

Ocultar thumbs Ver también para Turbo-V70:
Tabla de contenido
Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 19
− Disconnect the potentiometer, set it to 10 KΩ, and
then reconnect it.
− Connect the power cable.
− On the controller front panel push the START and
CURRENT pushbuttons and check the front panel
display as shown in the following table.
During start-up
without
Soft Start mode
With
48Ω
dumm
y load
Current ±10%
1.4 A
Power ±10%
80 W
Speed
34
±4 KRPM
52 °C
Temperature
±2 °C
− Check the 3-phase ac output voltage.
After start up it should be:
42 Vac with 78Ω dummy load and
42 Vac with 48Ω dummy load.
On the controller front panel, press the STOP
pushbutton and remove the power cable.
After
start-up
With
With
With
78Ω
48Ω
78Ω
dumm
dumm
dumm
y load
y load
y load
0.6 A
1.3 A
0.45 A
40 W
70 W
24 W
75
9
75
52 °C
52 °C
52 °C

Pump over-Temperature Test

− Disconnect the potentiometer, set it to 5KΩ, and
then reconnect it.
− Disconnect the dummy loads.
− Connect the power cable.
− On the controller front panel push the START
pushbutton.
The display should be as shown in the previous
figure and by pressing the CURRENT pushbutton
the temperature reading should be 71 °C ±2 °C. All
the other values should be zero.
− Check
the
It should be zero.
− On the controller front panel, press STOP and
remove the power cable.

Functional Test

Perform the functional test with the turbopump, taking
care to check the ramp sequence and start up time.
68
TECHNICAL INFORMATION
3-phase
ac
output
87-900-869-01 (B)
voltage.
Tabla de contenido
loading

Este manual también es adecuado para:

969-9507

Tabla de contenido