Greenlee CMT-60 Manual De Instrucciones página 10

Tabla de contenido
Idiomas disponibles

Idiomas disponibles

All manuals and user guides at all-guides.com
Accuracy
See the Specifications section for operating conditions and temperature
coefficient.
Accuracy is specified as follows: ± (a percentage of the reading + a fixed
amount) at 23 °C ± 5 °C (73.4 °F ± 9 °F), 0% to 80% relative humidity.
Accuracy Table
Characteristic
AC Current
0 to 200.0 A
AC Voltage*
0 to 600 V
DC Voltage*
0 to 600 V
0 to 2000 Ω
Resistance**
* Input impedance: 1.6 MΩ, 100 pF nominal
** Open circuit voltage: 3 V nominal
Continuity
Tone on: Circuit resistance is ≤ 25 Ω (approximately).
Tone off: Circuit resistance is ≥ 400 Ω (approximately).
Specifications
Display: 3-1/2-digit LCD
Maximum Conductor Size: 22 mm (0.87")
Sampling Rate: 2.5 per second
Automatic Power Off: After approximately 30 minutes
Temperature Coefficient: 0.2 x (Accuracy) per °C below 18 °C or above 28 °C
Overvoltage Protection Categories:
V-Ω Terminal: Category III, 600 VAC and 600 VDC
COM Terminal: Category III, 600 VAC and 600 VDC
Jaws: Category III, 600 VAC and 600 VDC
Operating Conditions:
0 °C to 30 °C (32 °F to 86 °F), 0% to 80% relative humidity
30 °C to 40 °C (86 °F to 104 °F), 0% to 75% relative humidity
40 °C to 50 °C (104 °F to 122 °F), 0% to 45% relative humidity
Indoor use only
Storage Conditions: –20 °C to 60 °C (–4 °F to 140 °F),
0% to 80% relative humidity
Remove battery
Elevation: 2000 m (6500') maximum
Pollution Degree: 2
Battery: 9 V battery (NEDA 1604, JIS 006P or IEC 6F22)
10
Range
Accuracy
± (2% + 0.3 A)
± (1.5% + 3 V)
± (1% + 2 V)
± (1% + 2 Ω)
Frequency Range
50 to 60 Hz
40 to 500 Hz
N/A
N/A
Tabla de contenido
loading

Tabla de contenido