The Pulse Oximeter is intended for use in the electromagnetic environment specified below. The customer or the user of the device
should assure that it is used in such an environment.
Immunity
IEC 60601-
test 
test level 
3 V/m80
Radiated RF
MHz bis 2,5
IEC 61000-4-3
GHz
Electromagnetic immunity 
Compliance
Electromagnetic environment–guidance 
level 
Portable and mobile RF communications  equipment should be used no closer to any  part of the
thermometer, including cables  than the recommended separation distance  calculated from the
equation applicable to the frequency of the transmitter.Recommended separation distance
d=1.2 √P
d=1.2 √P 80 MHz to 800 MHz
d=2.3 √P 800 MHz to 2,5 GHz
where P is the maximum output power rating of the transmitter in watts (W) according to the
transmitter manufacturer and d is the recommended separation distance in meters(m). Field
3 V/m
strengths from fixed RF transmitters, as determined by an electromagnetic site survey, should be
less than the compliance level in each frequency range. Interference may occur in the vicinity of
equipment.
43 