Calibration Mode

GB
Proceed to the next value to be edited by pressing [!] key.
!
After setting the minutes and pressing the [!] key the display will show
"IS SET" and the instrument returns to the measurement mode.
Store
Date
Cal
4
Time
Cal
Note:
cAL
user calibration (Display in calibration mode)
CAL
factory calibration (Display in calibration mode)
CAL
After confirming the selection with the [MODE] key the instrument will
show CAL/LR.
LR
Scroll through ranges using the [MODE] key.
Place the blank in the adapter making sure that the marks are aligned.
Press the [ZERO/TEST] key.
Zero
Test
The "Range" symbol flashes for approx. 8 seconds.
RANGE
The display shows the following in alternating mode:
0.0.0
CAL
Perform measurement with a standard of known concentration as
described.
Press the [ZERO/TEST] key.
Zero
Test
The "Range" symbol flashes for approx. 3 seconds.
RANGE
The result is shown in the display, alternating with CAL.
RESULT
If the reading corresponds with the value of the calibration standard
CAL
(within the specified tolerance), exit calibration mode by pressing the
[ON/OFF] key.
Changing the displayed value:
Pressing the [MODE] key once increases the displayed value by 1 digit.
Mode
Pressing the [ZERO/TEST] key once decreases the displayed value by
Zero
1 digit.
Test
Press the corresponding key until the reading equals the value of the
CAL
calibration standard.
RESULT + x
By pressing the [ON/OFF] key, the new correction factor is calculated
On
and stored in the user calibration software.
Off
Confirmation of calibration (3 seconds).
:
:
Note: The instrument calibration in the range MR is automatically taken also for HR.
24
All manuals and user guides at all-guides.com

User calibration

4
AL110_2 03/2018