5 User Modes
The Orion3 has 2 User Modes. Please note that these modes are only available if they
have been enabled by your practitioner.
Warning:
Care must be taken to only use those modes as instructed by your practitioner.
1 Flexion Lock
Sets the limit of flexion to a predetermined angle.
The maximum lock angle is 45 degrees from fully extended.
2 Cycling (Free swing)
Releases stance resistance, the Orion3 knee will then swing freely.
Warning:
In Cycling mode there is no stance control active, this mode must be de-activated
before resuming walking to avoid falling and possible injury.
Flexion Lock mode
User activation of Flexion Lock mode with
To activate Flexion Lock:
Fully extend the knee.
Press the
button 3 times, a short beep confirms each button press, 3 short beeps
confirm that Flexion Lock is ON.
To de-activate Flexion Lock:
Press and hold
you will hear 2 long beeps.
Cycling mode
Warning:
In Cycling (free swing) mode there is NO stance resistance.
User activation of Cycling mode with
To activate Cycling mode:
Press the
button 4 times, a short beep confirms each button press, 4 short beeps
confirm that Cycling mode is activated.
To de-activate Cycling mode:
Press and hold
you will hear 2 long beeps.
All manuals and user guides at all-guides.com
button on the Orion3 knee
button on the Orion3 knee
9
938375/3-1117