Sartorius YBT01 Instrucciones De Funcionamiento página 9

Radioenlaces para la comunicación inalámbrica
Tabla de contenido
Idiomas disponibles

Idiomas disponibles

Software from Sartorius
Sartorius offers two software programs
for fast and easy processing of the data
and control commands transmitted
between connected devices:
– WinScale
– SPEXXIS
Each of these programs comes with
complete instructions for simple instal-
lation on the PC used with your wireless
interfaces.
Configuring the Weighing Instrument
Data transmission parameters in the
RS-232 data interface on the YBT01 are
configured as follows:
– Transmission speed
(baud rate): 9600 bit/s
– Parity:
Odd
– Stop bits:
1
– Data bits:
7
Data transmission parameters in the
RS-232 data interface on the YDO01C-BT
(Option B2) are configured as follows:
– Transmission speed
(baud rate): 9600 bit/s
– Parity:
Odd
– Stop bits:
1
– Data bits:
8
Thus the data interface on the weighing
instrument must be adapted for data
transmission to the YBT01 or YDO01C-
BT (Option B2) wireless interface. Please
observe the following when configuring
the weighing instrument:
The weighing instrument models suit-
able for wireless data communication
use either the SBI (Sartorius Balance
Interface) or the XBPI (Xtended Balance
Protocol Interface) data transfer proto-
col. Which protocol is used in your
weighing instrument is indicated in the
operating instructions for the particular
instrument.
The XBPI protocol generally operates
with the following parameters:
– Transmission speed
(baud rate): 9600 bit/s
– Parity:
Odd
– Stop bits:
1
– Data bits:
8
With the SBI protocol, the transmission
speed, parity and number of stop bits
can be configured. The following
Sartorius instruments also permit con-
figuration of the number of data bits:
– GENIUS (ME models)
– MA100 and MA50 (electronic moisture
analyzers)
– Combics (CISL... models) (indicators and
complete scales)
– Combics (CIS... models) (indicators and
complete scales)
Refer to the installation and operating
instructions for the particular instru-
ment for details on which parameters
can be configured.
To enable communication with the
YBT01 or YDO01C-BT (Option B2) wire-
less interface, the data interface on the
weighing instrument must be config-
ured accordingly.
To use the SBI protocol and a YBT01
with its factory-set parameters, config-
ure the weighing instrument as follows:
– Transmission speed
(baud rate): 9600 bit/s
– Parity:
Odd
– Stop bits:
1
– Data bits:
7
9
Tabla de contenido
loading

Este manual también es adecuado para:

Ydo01c-bt

Tabla de contenido