Frequency; Resistance; Diode Test; Continuity Check - Bahco BDACM1000 Manual De Usuario

Ocultar thumbs Ver también para BDACM1000:
Tabla de contenido
Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 30
ACV:
Range
Resolution
4V
1mV
40V
10mV
400V
100mV
1000V
1V
Above accuracies can be guaranteed within 5%~100% of the full range.
The true RMS meter has residual value within 10 counts when the test leads are
shorten, but that will not affect the accuracy of measurement.
1. Frequency Range for ACV: 40Hz~1000Hz
2. Response for ACV: RMS measure, calibrated in rms of sine wave.
3. Overload Protection: 1000V DC or 1000V AC RMS.
4. Input Impedance (Nominal): DC voltage: >10MΩ; AC voltage: >10MΩ.

4.2.2 Frequency

Logic frequency (1Hz-1MHz)
Range
Resolution
4Hz
0.001Hz
40Hz
0.01Hz
400Hz
0.1Hz
4kHz
0.001kHz
40kHz
0.01kHz
400kHz
0.1kHz
4MHz
1kHz
Accuracy
±(1.0% of rdg + 5 digits)
±(1.5% of rdg + 5 digits)
Accuracy
±(0.1% of rdg + 5digits)
Voltage inpunt >10Vpp

4.2.3 Resistance

Range
Resolution
400Ω
0.1Ω
4kΩ
40kΩ
10Ω
400kΩ
100Ω
4MΩ
1kΩ
40MΩ
10kΩ

4.2.4 Diode Test

Range
Resolution
2000 V
0.001V

4.2.5 Continuity Check

Range
Resolution
200Ω
0.1Ω
Description: Continuity beeper ≤50Ω

4.2.6 Capacitance

Range
Resolution
4nF
1pF
40nF
10pF
400nF
100pF
4µF
1nF
40µF
10nF
400µF
100nF
4mF
1uF
Accuracy
±(0.5% of rdg + 3 digits)
±(0.5% of rdg + 2 digits)
±(1.5% of rdg + 5 digits)
Test Condition
Forward DC current approximately 1mA.
Reversed DC voltage approximately 1.5V.
Test Condition
Open circuit voltage: approx. 0.5V
Accuracy
±(5.0% of rdg + 20 digits)
±(3.0% of rdg + 15 digits)
±(5.0% of rdg + 10 digits)
EN
13
Tabla de contenido
loading

Tabla de contenido