Leveling the Balance
(only for models BBL31, BBL32, BBL42 or BBL52)
Level the balance any time you
set it up in a new location.
Use only the 2 front feet of the
balance for leveling.
§ Turn the 2 rear feet until they
§ Turn the 2 front feet as shown
> In most cases this will require
Anti-theft Locking Device
To protect against theft, use the
mounting lug on the rear panel of
the balance.
§ Secure the balance at the
Operation
Basic Weighing Function
Available Features
– Taring the balance
You can tare the balance within the entire weighing range.
Preparation
§ Turn on the balance: Press [ON/OFF]
§ To change configurations: See the chapter entitled
"Configuration"
§ To load factory-set configurations:
See "Configuration," parameter 9 – 1
$ To tare the balance: Press t
Additional Functions
§ To turn off the balance: Press [ON/OFF]
Warmup Time
To deliver exact results, the balance must warm up for
at least 30 minutes after initial connection to AC power. Only
after this time will the balance have reached the required
operating temperature.
All manuals and user guides at all-guides.com
are in position
(only on model BBL52)
here in the illustration until the
air bubble is centered in the
level indicator
several adjustment steps
place of installation,
for example with a chain
or a lock
Example
Basic weighing
Menu code settings: Factory-set codes
Step
Key (or instruction) Display
1. Turn on the balance
[ON/OFF]
Self-test is performed
2. Place container on balance
(here: 52 g)
t
3. Tare the balance
4. Place sample in container
on balance (here: 150.2 g)

Calibration/Adjustment*

Always calibrate/adjust the balance after setting it up in
a new location.
Available Features
Calibration/adjustment can only be performed when
– there is no load on the balance,
– the balance is tared, and
– the internal signal is stable.
If these conditions are not met, an error message
is displayed. Otherwise, the weight required for calibration/
adjustment is displayed.
Calibration/adjustment can be performed with any of three
weight units: g, kg
1
), or lb (menu code
Factory setting: Grams (menu code
1
) not for models with a readability of 0.1 mg
Example
Calibrate the balance
Menu code settings: Factory-set codes
Step
Key (or instruction) Display
1. Turn on the balance
[ON/OFF]
t
2. Tare the balance
3. Begin calibration
[CAL/CF]
Calibration weight is
displayed without
weight unit (here: 1000 g)
4. Place the indicated
calibration weight
on the balance
After calibration, the
calibration weight
is displayed with wt. unit
5. Remove the
calibration weight
* "Calibration" technically means to determine the difference between the
balance readout and the actual weight on the pan to determine the accuracy.
Adjustment means to bring a balance into the state of accuracy required for
its use. Therefore, "calibration," as used in this manual, actually means
"adjustment."
52.0
+
g
0.0
+
g
150.2
+
g
1. 4.
x)
1. 4. 1
)
0.0
g
1000.0
+
1000.0
1000.0
+
g
0.0
g
5