Symptom
The lifting column is wobbly
The lifter moves unevenly on a plane surface
The lifting arm moves uncontrolled.
The lifting legs move uncontrolled.
The lifter does not respond to hand control action
/ The lifting arm does not move.
Battery does not charge
Troubleshooting
Possible cause / action
The lifting column is not correctly positioned in
the chassis.
Assemble column and chassis, and lock column by
the locking handle on the chassis.
The castors move unevenly because of pile and
dust.
Clean the castors. Remove hair and dirt.
Short circuit between hand control and lifter.
Clean hand control plug surfaced with alcohol to
remove grease.
Defective hand control.
Replace hand control.
The control electronics has failed.
Contact local representative for service.
The leg opening mechanism has failed. The
product is unsafe to operate . Contact local
representative for service
Emergency stop button is activated.
Turn button clockwise to reset emergency stop.
Battery is empty.
Change or charge battery.
The lifters electronics is overheated.
Wait for it to cool down.
The lifter is overloaded.
Make sure the load does not exceed SWL.
Hand control is not plugged in properly.
Plug in Hand control properly. Hand control, plug
or cord can be broken an should be replaced.
Defective control electronics.
Contact local representative for service.
Faulty battery.
Check indicators on charger when charging.
Try another battery.
Faulty battery charger.
Check indicators on charger when charging. See
chapter "Charging".
Try another charger.