All manuals and user guides at all-guides.com
WARNING!
These short instructions are intended for commissioning the
Stratos Eco 2405 Cond transmitters. They are intended for
experienced users and do not replace the instruction manual
(on included CD-ROM or at www.knick.de).
The shipment includes a special document with safety
information.
Stratos Eco 2405 Cond is used for measurement of electrical conduc-
tivity and temperature in liquids. Fields of application are: biotechnol-
ogy, chemical industry, environment, food processing, water/waste-
water treatment.
The sturdy molded enclosure can be fi xed into a control panel or
mounted on a wall or at a post.
The protective hood provides additional protection against direct
weather exposure and mechanical damage.
The device can be used with all 2- and 4-electrode sensors. It provides
two current outputs (for transmission of measured value and tem-
perature, for example), two contacts, and a universal power supply
24 ... 230 V AC/DC, AC: 45 ... 65 Hz.
Terminals:
Screw terminal, suitable for single wires / fl exible leads up to 2.5 mm
(AWG 14).
Recommended torque for the terminal screws: 0.5...0.6 Nm.
Caution!
Additional safety precautions have to be taken for operation in
hazardous locations according to CSA (CLI, DIV2, GPA,B,C,D T4,
Ex nA IIC T4) (see
special document)!
Intended Use
2
3