Error Messages; Error Display - Knick STRATOS Eco 2405 Cond Guia De Inicio Rapido

Ocultar thumbs Ver también para STRATOS Eco 2405 Cond:
Tabla de contenido
Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 61

Error Messages

All manuals and user guides at all-guides.com
Error
Display
ERR 01
Measured
value
blinks
ERR 02
Measured
value
blinks
ERR 98
"Conf"
blinks
ERR 99
"FAIL"
blinks
12
Problem
Possible causes
Sensor
• Wrong cell constant
• Measurement range violation
• SAL > 45 ‰
• Sensor connection or cable defective
Unsuitable sensor
Conductance range > 3500 mS
System error
Configuration or calibration data defec-
tive; completely reconfigure the device
using the factory settings. Then calibrate.
Memory error in device program
Factory settings
EEPROM or RAM defective
This error message only occurs in the
case of a total defect. The device must
be repaired and recalibrated at the
factory.
x
x
x
x
x
x
x
x
x
x
x
x
x
x
Tabla de contenido
loading

Tabla de contenido