RELΔ
• Relative mode for Current measurement.
• Select manual range for Voltage and Resistance measurement.
• Change between Hz and duty cycle %.
Hold
• ON/OFF for hold function. H shows on display when value is
frozen.
Voltage measurement DC and AC
1. Insert red test lead into the HzDuty%VΩ terminal and black test lead into the
COM terminal.
2. Set the rotary switch to V position.
3. Select DC or AC with select button. Display shows DC or AC.
4. Push RELΔ button for manual ranging. AUTO disappere from the display. When
the value is
unknown always start from highest range.
5. Connect the test leads across with the object being measured. The measured
value shows on the display.
Current measurement DC and AC
1. Set the rotary switch to 40 A or 400 A position. Start with 400 A when the value
is unknown.
2. Select DC or AC with select button. Display shows DC or AC.
(See fig 1)
(See fig 3).