5 Changing the battery
The Li-ion mPack LL only needs replacement when the battery no longer
recharges to give sufficient capacity. This usually occurs after 3 to 5 years
if properly used.
• Disconnect the plug-in transformer of the mPack LL from the mains or
remove the mPack from the base unit (7).
• Unscrew the lid screw on the back of the mPack LL and remove the lid.
• Remove the old battery and disconnect the cord. No tools are necessary.
• Insert the new Li-ion mPack LL battery into the mPack so that the label is
visible and the battery cord is directed to the front control (3).
• Connect the plug of the battery to the printed circuit board.
• Take care when closing the lid that the cord is not pinched between lid and
body.
• To close, locate the two pins on the lid in the cut-outs in the lower half of the
housing. Close the lid and tighten the screw.
• Charge the new Li-ion mPack LL battery for two hours before use.
Note: The HEINE Li-ion mPack LL
Care and maintenance
• Always disconnect from the mains before cleaning.
• We recommend that external surfaces of the mPack LL be wiped clean with
a disinfectant cleaning wipe. This should only be carried out by qualified
personnel.
• Any disinfectant used must be approved by the disinfectant manufacturer
for use on plastics. The manufacturer`s instructions must be followed.
• The external surfaces may also be wiped clean with a cloth moistened with
water or with a neutral pH or slightly alkaline cleansing agent.
Isopropyl alcohol may only be used at concentrations up to 70%.
• Spray disinfection and soaking are not permitted and may cause damage.
• The article can be neither machine-cleaned, nor disinfected at high
temperature or sterilised. These processes lead to irreparable damage to
the product.
• Correct operation of the device must be tested at appropriate intervals with
regard to operating conditions.
• Use the units only when they are functioning correctly.
Specifications
HEINE base unit:
Mains voltage:
Safety class:
Current consumption:
Output:
Fuse:
Duration of use:
Length of cord:
HEINE mPack LL
Output:
Input:
Fuse:
Charging current limit:
Weight:
Dimensions:
®
100– 240 V~ / 50 – 60 Hz
Double insulated
max. 30 VA
9 V DC max. 3 A
Integral overload protection
Continuous
200 cm
:
®
Variable, max. 7,5 W
9 V DC, max. 3 A DC
Integral overload protection
3 A
ca. 300 g
80 mm x 28 mm x 125 mm
battery must be disposed of properly.