Filter Cartridge Replacement Procedure

1
2
Exhaust Cap
Figure 26 Filter Cartridge Replacement
3
202
Unscrew the cap of the oil drain valve and screw the oil drain pipe to drain the
oil. Let the oil flow out. In the meantime, clean the inside of the drain plug.
After closing the drainage hole with its plug make the pump perform a number
of turns, giving it a brief current pulse, so as to let residual oil in the pump
chambers flow out. Then drain off the residual oil again.
After the oil has been drained, close the drain plug tight; take the tank of waste
oil away and clean thoroughly, using rags of cotton or other suitable material to
dry.
Open the fill oil cap using a 10 mm hex key and add AVF Platinum oil until
reaching the maximum level on the sight glass.
Close the oil fill cap tight.
Replace the cartridge filter as per following slide.
Unscrew the cap rotating it anti-clockwise, by hand or socket wrench with size
19 mm.
Pull out the cartridge filter.
Insert new cartridge filter and double check it is in the right position (inserted in
the appropriate hole).
Cartridge Filter P/N 9499201
MS40-S User Manual