C A L I B R AT I O N
Regular calibration of the probes is important.
The calibration is carried out by means
of the reference solution supplied.
Clean the SALINITY MANAGER with the
Cleaning Solution for about 2 minutes.
Then clean the SALINITY MANAGER in
osmosis water and shake any residual water
out of the device.
Unscrew the upper lid and immerse the
SALINITY MANAGER in the calibration solu-
tion. If necessary, adjust the measured value
to 35 ppt with the included screwdriver.
Further technical information and operating
aids can be found in our
HTU ( How To Use ) SALINITY MANAGER
For download at
www.faunamarin.de DownloadCenter
1.
2.
35 ppt
Cleaning
osmosis
Solution
water
(2 min)
SALINITY MANAGER
calibration solution
E N