GB
9. HIGH limit alarm setup
9.1 Press Mode key 5 times or until H symbol is flashing to select HIGH limit alarm setup (shown below).
9.2 Press UP or DOWN arrow key to turn ON and OFF the function.
9.3 When in ON position, press MODE key to see the set alarm value.
9.4 Use UP or DOWN arrow key to change the value.
Note: The value must be higher than LOW limit alarm. Pressing the key shortly will result in value changing
1 digit at the time and holding down the key down will result in counting continually. Instrument will
exit setup after 5 seconds if there are no operations.
10. Low limit alarm setup
10.1 Press Mode key 7 times or until L symbol is flashing to select Low limit alarm setup (shown below).
10.2 Press UP or DOWN arrow key to turn ON and OFF the function.
10.3 When in ON position, press MODE key to see the set alarm value.
10.4 Use UP or DOWN arrow key to change the value.
Note: The value must be lower than HIGH limit alarm. Pressing the key shortly will result in value changing
1 digit at the time and holding down the key down will result in counting continually.
Instrument will exit setup after 5 seconds if there are no operations.
11. Find heat and cold point
ter at target area and start to measure, then move up and down slowly to sweep the whole area until the heat
point and/or cold point are found.
12. D:S (Distance Coefficient)
Limit 96 has D:S of 10:1 which means that if an object is measured from a distance of 10m the instrument
will show the average value of 1m in diameter (shown below).
25@250 50@500
1@10
2@20
3@30
75@750
D:S = 10:1
4
English