Max/Min Function - LIMIT 96 Manual Del Usuario

Tabla de contenido
Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 59
Measuring unit (6)
Trigger free setup (8)
High alarm setup (9)
High alarm value (9)
Low alarm setup (10)
Low alarm value (10)
5. Emissivity setup
5.1 Press Mode key 1 times or until
5.2 Press UP or DOWN arrow key to select desired value according to Emissivity values table.
Note: Emissivity can be adjusted from 0.01 to 1.0. Pressing the key shortly will result in value changing 1
digit at the time and holding down the key down will result in counting continually.
Instrument will exit setup after 5 seconds if there are no operations.
6. Unit setup
6.1 Press Mode key 2 times or until C or F start flashing to enter Temperature unit setup (shown below).
6.2 Press UP or DOWN arrow keys change between Celsius (°C) and Fahrenheit (°F).
Note: Instrument will exit setup after 5 seconds if there are no operations.

7. MAX/MIN function

7.1 Press MODE 3 time to select MAX/MIN function (shown below).
7.2 Press arrow UP or DOWN key to change between the different functions.
7.3 The value of selected function will be shown in the secondary display.
MAX = shows the highest value in the secondary display.
MIN = shows the lowest value in the secondary display.
8. Trigger Free Measurement
8.1 Press Mode key 4 times to enter trigger free measurement setup (shown below) or until
flashing on the LCD screen.
8.2 Press UP or DOWN arrow key to turn ON or OFF the trigger free function. Instrument will exit setup
after 5 seconds if there are no operations.
8.3 When trigger free function is selected the thermometer will continuously measure until you press the
trigger.
8.4 Press the trigger again,
measuring has stopped, and the last value will be saved in the display.
Note: The target must be bigger than the diameter of the two laser spots to ensure accurate reading. Please
also make sure to select correct Emissivity value.
EMS Setup (5)
MIN/MAX (7)
symbol is flashing to enter emissivity setup (shown below).
and SCAN icon will disappear, and HOLD icon appear to indicate that the
Change value
Select °C or °F
Select Min, Max, function
Select On or OFF
Select On or OFF
Change value
Select On or OFF
Change value
3
English
GB
icon is
Tabla de contenido
loading

Este manual también es adecuado para:

95