Power Failure

In the event of a power failure (momentary or long term), the Turbo-
V controller will stop the turbopump and all the interconnected
pumps/devices. The Turbo-V vent valve device, if used, will vent the
turbopump only if the power failure is longer than the preset delay
time. When power is restored, the Turbo-V controller automatically
restarts the interconnected devices and the turbopump in the proper
sequence.
The display shows:
P
1
until normal operation achieved.

Remote Control Mode Operation

If remote signals are used to operate the controller, it must be
programmed for remote operation (see paragraph
"FRONT/REMOTE/232 Selection") and when ready to start, the
display shows as in the following figure.
R
S
If the Soft Start has been deselected the display shows:
P
With or wihout Soft Start mode selected the START/STOP and LOW
SPEED pushbuttons are inoperative, while the CYCLE NUMBER and
PUMP CURRENT pushbuttons are always active.
Turbo-V 70 PCB Controller User Manual / 87-900-871-01 (E)