Operation; Start Up - KNF SC 920G Manual De Instrucciones

Tabla de contenido
Idiomas disponibles
  • ES

Idiomas disponibles

Vacuum Pump System SC 920G
Translation of original Operating Instructions, English, KNF 334631-334633 07/21
7.

Operation

7.1. Start up

Before switching on the vacuum pump system, observe the
following points:
Operational requirements
All hoses attached properly
Fan openings not blocked
Specifications of the power supply correspond with the data
on the vacuum pump system's type plate.
Recirculating cooler or cold water connection ready on high-
performance condenser.
The high-performance condenser's gas outlet is not blocked
(high-performance condenser is not pressure-proof).
Vacuum pump system is at room temperature.
No reactive, explosive or otherwise hazardous mixtures may
be produced when ventilating the vacuum system through the
air inlet (Fig. 2/6, page 11) (if necessary, use an inert gas)
When the gas ballast valve is open, no reactive, explosive, or
otherwise hazardous mixtures can form (otherwise, contact
KNF Service).
Tab. 3
 Only operate the vacuum pump system under the operating
parameters and conditions described in chapter 4, Technical
data (page 8).
 Make sure the vacuum pump system is used properly (see
chapter 2.1, page 5).
 Make sure the vacuum pump system is not used improperly
(see chapter 2.2, page 5).
 Observe the safety precautions (see chapter 3, page 6).
Personal injury caused by poisoning or explosion
and damage to the vacuum pump system.
 Make sure that no reactive or explosive mixtures
will be produced when ventilating the vacuum
WARNING
pump system through the air inlet.
Operation
DE
EN
FR
ES
IT
19
Tabla de contenido

Solución de problemas

loading

Tabla de contenido