- Check that there is no humidity in the
handpiece cord connections. If there is, dry it
out by wiping, then blowing with the
multifunction syringe.
- Connect the handpiece to the cord connector by
simply plugging it in, making sure that the two
marks on the cord plug and the handpiece are
lined up.
- Place the handpiece on its rest.
- Install the irrigation solution tanks on their
supports.
- Place the covers on the irrigation solution
tanks.

4. 4 FIRST USE

Before using the P M
time, you are advised to carry out the following
operations:
- Fill the irrigation solution tanks with
disinfectant, without exceeding the level
indicated by the borders (Fig. 4-1).
- Place the covers on the irrigation solution tanks.
- Place the irrigation solution tanks on the
device.
- Connect the handpiece cord to the device.
- Connect the handpiece to the handpiece cord.
- Switch on the device (I).
- Set the device to minimum power (1) and
maximum flow.
- Select an irrigation solution tank using the push
button (Fig. 1-5).
10
N
XS for the first
AX
EWTRON
- Press the irrigation flow setting button (Fig. 1-
6) to rinse the irrigation circuit on the device.
- Completely empty the disinfectant solution out
of the tank.
- Select the other irrigation solution tank by
pressing the tank selection push button again
(Fig. 1-5).
- Press the irrigation flow setting button (Fig. 1-
6) to rinse the irrigation circuit on the device.
- Completely empty the disinfectant solution out
of the tank.
- Remove the irrigation solution tanks from the
device.
- Remove the covers from the irrigation solution
tanks.
- Fill the irrigation solution tanks with distilled or
demineralized water.
- Place the covers on the irrigation solution
tanks.
- Place the irrigation solution tanks on the
device.
- Select an irrigation solution tank using the push
button (Fig. 1-5).
- Press the irrigation flow setting button (Fig. 1-
6) to rinse the irrigation circuit on the device.
- Completely empty the tank.
- Select the other irrigation solution tank by
pressing the tank selection push button again
(Fig. 1-5).
- Press the irrigation flow setting button (Fig. 1-
6) to rinse the irrigation circuit on the device.
- Completely empty the tank.
- Select the air function by pressing the
corresponding switch on the foot pedal (Fig. 3-
4).
- Flush the irrigation circuit with air for at least
30 seconds by activating the ultrasonics using
the control footswitch (Fig. 3-2).