Pro-care 2 / English

2.3 Front panel

1. Pressure Adjustment Knob
Pressure adjustment knob controls the air pressure output. When turning clockwise, the output pres-
sure will increase. Vice versa for decreasing air pressure. Higher pressure output will support the heavi-
er weight patient. The setting reference is given as Kg. Please consult your physician for a suitable set-
ting.
NOTE- Check if the pressure is suitable for the patient by sliding one hand beneath the air cells at
the level of the patient's buttocks. Always leave at least 1" inch space between patient and the
static cell to prevent bottoming out.
2. Normal Pressure Indicator
When green LED lights up, the pressure inside of air mattress reaches the preset pressure setting.
3. Low Pressure Indicator
When low pressure LED lights up, the pressure inside of air mattress is below normal. Please refer to
troubleshooting. During initial inflation of the mattress, the Low Pressure LED will illuminate until the
set pressure is reached.
4. Static Function
Press blue STATIC button to suspend alternating function, if needed. The pressure inside of air tubes
will be adjusted to the same softness. Press the STATIC button again; it will switch back to alternating
mode.
5. Alarm Mute
Press alarm mute button to temporary suspend the low-pressure alarm. If the situation did not resolve
in 3~5 mins, the alarm will resume again to notify the caregiver.
6. Power Failure Alarm (PFA)
During power failure situation, the PFA LED light and buzzer will notify caregivers to take appropriate
action. By pressing the alarm mute button both PFA buzzer and LED will cease.