Pro-care 2 / English
Guidance and Manufacturer's Declaration- Electromagnetic Immunity:
This device is intended for use in the electromagnetic environment specified below. The user of this device
should make sure it is used in such an environment.
Immunity Test Levels
Basic EMC stand-
Professional
ard
healthcare
facility environment
Electrostatic
±8kV contact
Discharge (ESD)
±15kV air
IEC61000-4-2
Electrical fast
±2kV for power supply line
transient/ burst
±1kV for input/output line
IEC61000-4-4
± 1 kV line(s) to
Surge
line(s)
IEC61000-4-5
± 2 kV line(s) to
earth
Voltage dips,
Voltage Dips:
short interrup-
i) 100% reduction for 0.5 period,
tions and voltage
ii) 100% reduction for 1 period,
variations on
iii) 30% reduction for 25/30 period,
power supply
Voltage Interruptions:
input lines
100% reduction for 250/300 period
IEC61000-4-11
Power frequency
(50/60Hz) mag-
30 A/m
netic field
IEC61000-4-8
3 Vrms
0,15 MHz – 80 MHz
Conducted RF
6 Vrms in ISM bands
IEC 61000-4-6
between 0,15 MHz
and 80 MHz
80 % AM at 1 kHz
3 V/m 80 MHz to
2.7 GHz
80 % AM at 1 kHz
Radiated RF EM
385-6000 MHz,
Fields
9-28V/m, 80%
IEC61000-4-3
AM(1kHz) pulse
mode and other
modulation
Compliance
HOME
Levels
HEALTHCARE
ENVIRONMENT
±8kV contact
±15kV air
±2kV for power
supply line
±1kV for in-
put/output line
± 1 kV line(s) to
± 1 kV line(s) to
line(s)
line(s)
230 V
30 A/m
30 A/m
3 Vrms
0,15 MHz – 80
MHz
6 Vrms in ISM and
amateur radio
6Vrms
bands between
0,15 MHz and 80
MHz
80 % AM at 1 kHz
10 V/m 80 MHz to
2,7 GHz
80 % AM at 1 kHz
385-6000 MHz,
10V/m
9-28V/m, 80%
AM(1kHz) pulse
mode and other
modulation
14
Electromagnetic Environ-
ment-Guidance
Floors should be wood, con-
crete or ceramic tile. If floors
are covered with synthetic
material, the relative humidity
should be at least 30 %.
Mains power quality should be
that of atypical commercial or
hospital environment
Mains power quality should be
that of atypical commercial or
hospital environment.
Mains power quality should be
that of a typical commercial or
hospital environment. If the
user of this device requires
continued operation during
power mains interruptions, it
is recommended that the
device be powered from an
uninterruptible power supply
or a battery.
Power frequency magnetic
fields should be at levels
characteristic of a typical
location in a typical commer-
cial or hospital environment.
Portable and mobile RF com-
munications equipment
should be used no closer to
any part of this device, in-
cluding cables, than there
commended separation dis-
tance calculated from the
equation applicable to the
frequency of the transmitter.
Recommended separation
distance
d
=
150kHz to 80MHz
d =0.6
80MHz to 800MHz
d =1.2
800 MHz to 2.7GHz