0x10A5
Parameter setting- Alarm setting
0x10A5
4261 PFC_ALARM_SET_MAX_FLOW_RATE-High
0x10A6
4262 PFC_ALARM_SET_MAX_FLOW_RATE+1
0x10A7
4263 PFC_ALARM_SET_MAX_FLOW_RATE+2
0x10A8
4264 PFC_ALARM_SET_MAX_FLOW_RATE-Low
0x10A9
4265 PFC_ALARM_SET_MIN_FLOW_RATE-High
0x10AA
4266 PFC_ALARM_SET_MIN_FLOW_RATE+1
0x10AB
4267 PFC_ALARM_SET_MIN_FLOW_RATE+2
0x10AC
4268 PFC_ALARM_SET_MIN_FLOW_RATE-Low
0x10AD
4269 PFC_ALARM_SET_EMPTY_TUBE_DETECT
0x10AE
4270 PFC_ALARM_SET_ALARM_1_FUNC
0x10AF
4271 PFC_ALARM_SET_ALARM_1_TYPE
0x10B0
4272 PFC_ALARM_SET_ALARM_2_FUNC
0x10B1
4273 PFC_ALARM_SET_ALARM_2_TYPE
0x10B2
4274 PFC_ALARM_SET_CURR_FUNC
0x10B3
4275 PFC_ALARM_SET_ALARM_CURR_VAL
MANUAL - MEDIDOR ELECTROMAGNÉTICO
Data
Unit
Range
Type
+0.0000 ~ Line
m³/s
FLOAT64
size(m)^2 *7.85
+0.0010~8.000
FLOAT64
m³/s
(Default:2.0)
0: Off
UINT16
N/A
1: On
0: None
1: Max. flow rate
UINT16
N/A
2: Min. flow rate
3: Empty alarm
4: System error
0:NO
UINT16
N/A
1:NO
0: None
1: Max. flow rate
UINT16
N/A
2: Min. flow rate
3: Empty alarm
4: System error
0:NO
UINT16
N/A
1:NO
0: none
UINT16
N/A
1: Empty alarm
2: System error
0:3.6
1:3.8
UINT16
mA
2:20.5
3:22
www.equysis.com
Definition
Authority
Read/
Write
Read /
Write
Max. flow rate
Read /
Write
Read /
Write
Read /
Write
Read /
Write
Min. flow rate
Read /
Write
Read /
Write
Empty pipe
Read /
detection
Write
Output 1 contact
Read /
function
Write
Output1 contact
Read /
form
Write
Output 2 contact
Read /
function
Write
Output 2 contact
Read /
form
Write
Alarm current
Read /
function
Write
Alarm current
Read /
setting
Write
45