INSTRUCTION MANUAL
Emergency stop
Machine
Safety free space
SWITCH
ON
OPERATING TIME:
ROTOR TIME:
NORMAL BUCKET TIME:
DOUBLE BUCKET TIME:
Accumulated working times of components
with limited service life
Main menu
J
.P. SELECTA s.a.u.
CODE 80343 REV A
6.1.3 Filling and location of the buckets
Once the buckets are filled, and before placing them in the
rotor, please check that their weight do not differ in more than
15 grams.
Do not start the centrifuge if the buckets are not placed.
And under no circumstances start the centrifuge with
the rotor loaded in an unbalanced way.
6.1.4 Emergency stop
If you have to stop the machine, quickly activate the front
panel button (18).
Note: Do not use this button as an usual form of stoppa-
ge. Please only activate it in case of emergency.
6.1.5 Safety measures for the staff
Do not lean on the centrifuge.
Do not stay in the safety free space more than the strictly
necessary.
Do not leave potentially hazardous materials inside the security
free space.
7 Operating menus
Press the ON/Off switch.
The hour counter of the centrifugal elements appears with
limited service life.
Rotor
Normal bucket.
00004h 10m
Double bucket.
00004h 10m
00004h 10m
After a few seconds, the main menu screen appears.
00004h 10m
You can select from the main menu, by pressing on the touch
screen:
SETTINGS: To update date, time, language, data download,
software updates, life counters initialization.
PROGRAMMING: To establish programs for samples
treatment:
WORKING: To select one of the stored programs.
Autovía A-2 Km 585.1 Abrera 08630 (Barcelona) España
- Times
- Speed
- Temperature.
09/2021
Page: 26
Tel 34 937 700 877 Fax 34 937 702 362