Operating Functions; Audible Alarm - Medline HEMO-FORCE MDS600INT2 Instrucciones De Operación

Ocultar thumbs Ver también para HEMO-FORCE MDS600INT2:
Tabla de contenido
Idiomas disponibles

Idiomas disponibles

OPERATING FUNCTIONS

1. BATTERY OPTION (Battery sold separately)
i.
This device has a Lithium battery option that can serve
as a backup. The device constantly monitors the outer AC
electric power supply. In the case of a power failure the
pump will switch to battery operation automatically.
ii.
The "Battery Indicator" indicates the battery
capacity status as illustrated.
iii. Under normal conditions a fully charged battery will last
approximately 4 hours of continuous operation. It may
take up to 16 hours to fully charge a depleted battery.
iv. The pump will show a flashing light from "Battery Indicator"
and trigger the audio alarm to give a low battery warning
5 minutes before the pump automatically shuts down.
2. INDICATOR LIGHT (behind power button)
i.
Green: Illuminated to indicate power on with AC power.
ii.
Amber: Illuminated to indicate power on with battery.
iii. Amber: Flashing to indicate a fault has been
detected followed by an audio alarm.

3. AUDIBLE ALARM

Press the alarm mute button and the audible alarm can be turned off.
The audible alarm will alarm after 10 minutes if the error still happens.