Clean the vial with the optional wiper
Carefully remove most of the water in the vial. Carefully put the vial wiper into the process vial so that no water
spills out.
Clean the process vial with the optional vial wiper as shown in the illustrated steps that follow.
Clean the vial compartment
Clean the vial compartment only when the compartment has contamination. Make sure that the tool
to clean the vial compartment has a soft surface and does not damage the instrument.
shows the options on how to clean the vial compartment.
Contaminant
Dust
Liquid, oil
Replace the vial
Keep water out of the vial compartment or instrument damage will occur. Before the automatic cleaning module is
installed on the instrument, make sure that there are no water leaks. Make sure that all tubing is fully seated.
Make sure that the vial nut is tight.
Hold the automatic cleaning module vertically when it is installed on the instrument or the vial can break. If the vial
breaks, water will get in the vial compartment and instrument damage will occur.
24 English
N O T I C E
Table 3 Cleaning options
Options
Vial compartment wiper, micro fiber cloth, lint-free cloth
Cloth, water and cleaning agent
N O T I C E
N O T I C E
Table 3