TO CHECK PLUMB
Choose a wall and set laser 5 m away from the wall. Hang a plumb with the
length 2.5 m on the wall. Turn on the line laser and make the vertical laser
line meet the point of the plumb. The accuracy of the line is in the range if
the vertical line doesn't exceed (up or down) the accuracy that is shown in
the specifications (±3 mm/10 m). If the accuracy isn't corresponding with
claimed accuracy, contact the authorized service center.
Note: Because of construction of laser emitter laser beam may be unho-
mogeneous and has different intensity of brightness along the perimeter
in different light conditions. Unhomogeneous of laser beam: laser patch of
light but the center of laser beam is identified. Different laser beam bright-
ness: difference of intensity is up to 50%.

CARE AND CLEANING

Please handle line laser with care. Clean with soft cloth only after any use.
If necessary damp cloth with some water. If instrument is wet clean and
dry it carefully. Pack it up only if it is perfectly dry. Transport in original
container/case only.
Note: During transport On/Off compensator lock (5) must be set to posi-
tion "OFF". Disregard may lead to damage of compensator.
SPECIFIC REASONS FOR ERRONEOUS MEASURING RESULTS
• Measurements through glass or plastic windows;
• Dirty laser emitting window;
• After line laser has been dropped or hit. Please check the accuracy;
• Large fluctuation of temperature: if instrument will beused in cold areas
after it has been stored in warm areas (or the other way round) please wait
some minutes before carrying out measurements.