B to another point C. Point C must be on the vertical laser line, it means
the height of C point is 3m. Measure the distance from point A to point
C, if the distance is over 2 mm, please, contact with seller to calibrate the
laser.
CARE AND CLEANING
Please handle measuring instrument with care. Clean with soft cloth only
after any use. If necessary damp cloth with some water. If instrument is
wet clean and dry it carefully. Pack it up only if it is perfectly dry. Trans-
port in original container/case only.
Note: During transport On/Off compensator lock (3) must be set to posi-
tion "OFF". Disregard may lead to damage of compensator.
SPECIFIC REASONS FOR ERRONEOUS MEASURING RESULTS
Measurements through glass or plastic windows;
Dirty laser emitting window;
After instrument has been dropped or hit. Please check the accuracy.
Large fluctuation of temperature: if instrument will be used in cold
areas after it has been stored in warm areas (or the other way round)
please wait some minutes before carrying out measurements.
ELECTROMAGNETIC ACCEPTABILITY (EMC)
It cannot be completely excluded that this instrument will disturb
other instruments (e.g. navigation systems);
will be disturbed by other instruments (e.g. intensive electromagnet-
ic radiation nearby industrial facilities or radio transmitters).
LASER CLASS 2 WARNING LABELS ON THE LASER INSTRUMENT
LASER CLASSIFICATION
The instrument is a laser class 2 laser product accortding to DIN IEC
60825-1:2007. It is allowed to use unit without further safety precautions.
SAFETY INSTRUCTIONS
Please follow up instructions given in operating manual.
Do not stare into beam. Laser beam can lead to eye injury (even from
greater distances).
Do not aim laser beam at persons or animals.
The laser plane should be set up so that the beam path is not at normal
eye level.